TB 9-6625-2252-35
(10) Set calibrator for a 100, 200 Hz output. Adjust A12R2 100V (fig. 1) until TI
indicates a null (R).
(11) Set calibrator for a 100, 20 kHz output. Adjust A13C2 100V (fig. 1) until TI
indicates a null (R).
(12) Set calibrator for a 1000, 200 Hz output. Adjust A12R3 1000V (fig. 1) until TI
indicates a null (R).
(13) Set calibrator for a 1000, 20 kHz output. Adjust A13C3 1000V (fig. 1) until TI
indicates a null (R).
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
9/(10 Blank)