TB 9-6625-2252-35
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Fluke, Model 5720A (5700A/EP)
CALIBRATOR
Dc volts:
(p/o MIS-35947) w/amplifier, Fluke,
Range: 1 to 100 V dc
Model 5725A/AR (5725A/AR)
Accuracy: .0125%
Ac volts:
Range: 1 to 1000 V ac
Accuracy: .05%
Range: -0.1 to 2.6 V dc
Fluke, Model 8840A/AF05
(AN/GSM-64D)
Accuracy: 1.0%
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in TM 11-6625-2724-12 for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.