TB 9-6625-2182-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
3165-14 was used as the prime data source in compiling these instructions. The equipment
being calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time
required for this
calibration is
approximately 6
2. Forms,
Records, and
Reports.
Forms, records, and
reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Range: 100 kHz to 2000 MHz
Accuracy: +10 PPM
Time stability: +0.05 PPM/hour1
Line stability: +0.05 PPM, 10% line change2
RF output
Range: >15 to -140 dBm3
Flatness:
+1.5 dB (100 kHz to 500 MHz)
+2 dB (500 to 1000 MHz)
+2.5 dB (1 to 2 GHz)
Attenuator accuracy: +2.0 dB
Spectral purity
Accuracy
100 kHz to 1057.5 MHz
<+13 dBm, <-25 dBc
1057.5 to 2000 MHz
<+7 dBm, <-25 dBc
Accuracy
100 kHz to 1057.5 MHz
<-100 dBc
1057.5 to 2000 MHz
<-45 dBc
Spurious signal range
Accuracy
100 kHz to 132.1875 MHz
<-70 dBc
132.1875 to 1057.5 MHz
<-90 dBc